Guidance for SPR and BLI binding characterization experiments. Use when: (1) Planning binding kinetics experiments, (2) Troubleshooting poor/no binding signal, (3) Interpreting kinetic data artifacts, (4) Choosing between SPR vs BLI platforms.
The English summary is being prepared. Showing the original description.
See how to installReview the instructions the agent will receive before you install.
| Factor | Choose SPR | Choose BLI |
|---|---|---|
| Sensitivity | Small molecules, fragments (<500 Da) | Large complexes, antibodies |
| Throughput | Low-medium (serial) | High (96-well parallel) |
| Sample purity | Required (clogs fluidics) | Tolerates crude lysates |
| Kinetic resolution | Higher (better for fast kinetics) | Lower |
| Mass transport | More sensitive (may distort kon) | Less sensitive |
| Maintenance | High (fluidics system) | Low (dip-and-read) |
| Sample consumption | Higher (continuous flow) | Lower |
| Cost per experiment | Lower chip cost, higher run cost | Higher tip cost, lower run cost |
| Cause | Mechanism | Solution |
|---|---|---|
| Hydrophobic CDRs | Adsorb to SPR gold/dextran surface | Add 0.05% Tween-20, use CM7 chip with longer dextran |
| Aggregation | Mass transport artifacts in SPR fluidics | Filter sample (0.22μm), reduce ligand density |
| High instability | Degrades during continuous flow | Shorter cycle time, add stabilizers (trehalose 5%) |
| Charge mismatch | Nonspecific binding to charged dextran | Adjust buffer pH ±1 from pI, add BSA 1mg/mL |
| Slow dissociation | Long regeneration needed (damages ligand) | Use BLI (disposable tips) |
| Cause | Mechanism | Solution |
|---|---|---|
| Small analyte | BLI less sensitive for <10 kDa | Use SPR with appropriate chip |
| Weak affinity (KD >10μM) | Fast dissociation in BLI dip | Increase analyte concentration |
| Low expression | Not enough signal | Increase biosensor loading |
Mass transport limitation occurs when analyte cannot diffuse to the surface fast enough to maintain equilibrium. This distorts kinetic parameters.
| Strategy | SPR | BLI |
|---|---|---|
| Reduce ligand density | <200 RU for high-affinity | <0.5 nm shift loading |
| Increase flow rate | 50-100 μL/min | Increase shake speed (1000 rpm) |
| Use oriented immobilization | His-tag capture | Biotinylated ligand |
| Include in fitting | Mass transport model (kt) | Usually less critical |
| Additive | Concentration | Mechanism | Best For |
|---|---|---|---|
| BSA | 0.5-1 mg/mL | Blocks hydrophobic sites | General use |
| Tween-20 | 0.02-0.05% | Prevents surface adsorption | Hydrophobic analytes |
| Trehalose | 1-5% | Stabilizes + blocks | Unstable proteins |
| Sucrose | 5% | BLI-specific blocker | BLI tips |
| Carboxymethyl dextran | 1 mg/mL | Competitive blocking | SPR with charged proteins |
| NaCl | 150-500 mM | Reduces ionic interactions | Charged proteins |
Always include:
| Condition | Targets | Caution |
|---|---|---|
| 10 mM Glycine pH 2.0-2.5 | Most protein-protein | May denature ligand |
| 10 mM Glycine pH 1.5 | Strong interactions | Harsh, limit exposure |
| 1-2 M NaCl | Ionic interactions | Mild, try first |
| 10 mM NaOH | Very stable ligands | Can hydrolyze proteins |
| 10 mM Glycine pH 9-10 | Acid-stable proteins | Can aggregate |
| 10 mM EDTA | His-tag, metal-dependent | Strips Ni-NTA |
| 4 M MgCl2 | Hydrophobic interactions | Check ligand stability |
Symptoms: Two-rate association or dissociation Causes:
Solutions:
Symptoms: Signal increases during dissociation phase Causes:
Solutions:
Symptoms: Signal decreases at high analyte concentrations Causes:
Solutions:
No reviews yet. Tell others what you thought.
No login needed. Share how it worked for you, ask a question, or note where it did not run.
No comments yet. Be the first to write one.
Workflow for read alignment, sorting, indexing, mapping statistics, and downstream-ready alignment artifacts.
The English summary is being prepared. Showing the original description.
Validate protein designs using AlphaFold2 structure prediction. Use this skill when: (1) Validating designed sequences fold correctly, (2) Predicting binder-target complex structures, (3) Calculating confidence metrics (pLDDT, pTM, ipTM), (4) Self-consistency validation of designs, (5) Multi-chain complex prediction with AlphaFold-Multimer. For faster single-chain prediction, use esm. For QC thresholds, use protein-qc.
The English summary is being prepared. Showing the original description.
AlphaFold2 / AlphaFold-Multimer structure prediction for validation and confidence scoring. Use this skill when: (1) Validating designed sequences fold correctly, (2) Predicting binder-target complex structures, (3) Calculating confidence metrics (pLDDT, pTM, ipTM), (4) Self-consistency validation of designs, (5) Multi-chain complex prediction with AlphaFold-Multimer. For faster single-chain prediction, use esm2-sequence-scoring. For QC thresholds, use protein-design-qc.
The English summary is being prepared. Showing the original description.
Workflow for event-level and isoform-level splicing analysis with sashimi-ready outputs and splice QC.
The English summary is being prepared. Showing the original description.
ATAC-seq processing with assay QC, MACS3 peak calling, consensus peak matrices, differential accessibility, and motif or footprint follow-up.
The English summary is being prepared. Showing the original description.
End-to-end binder design using BindCraft hallucination. Use this skill when: (1) Designing protein binders with built-in AF2 validation, (2) Running production-quality binder campaigns, (3) Using different design protocols (fast, default, slow), (4) Need joint backbone and sequence optimization, (5) Want high experimental success rate. For backbone-only generation, use rfdiffusion. For QC thresholds, use protein-qc. For tool selection guidance, use binder-design.
The English summary is being prepared. Showing the original description.