Corrects fluorophore spillover (conventional compensation) or spectral overlap (spectral unmixing) and applies variance-stabilizing transforms (logicle/biexponential, arcsinh, log) for flow and mass cytometry. Covers spillover-matrix estimation from single-stain controls, AutoSpill, the spillover spreading matrix and why panel design (not compensation) bounds resolution, compensate-then-transform ordering, and arcsinh cofactor choice (5 for CyTOF, ~150 for fluorescence, per-channel via flowVS). Use when correcting spectral overlap, preparing data for gating/clustering, choosing logicle vs arcsinh, deciding a cofactor, or distinguishing compensation from spectral unmixing.
インストールする前に、エージェントに与えられる指示の中身を確認できます。
Reference examples tested with: flowCore 2.14+, flowStats 4.14+, flowWorkspace 4.14+, CATALYST 1.26+.
Before using code patterns, verify installed versions match. If versions differ:
packageVersion('<pkg>') then ?function_name to verify parametersNotes that bite: estimateLogicle() lives in flowWorkspace (not flowCore). flowCore::spillover() on a flowFrame returns a LIST of keyword matrices (index [[1]]); flowStats::spillover() on single-stain controls returns the matrix DIRECTLY (not a list) - do not index it with $.
If code throws an error, introspect the installed package and adapt rather than retrying.
"Compensate and transform my cytometry data" -> Remove spillover (matrix subtraction, conventional) or unmix the full spectrum (least squares, spectral), then apply a transform so populations separate.
flowCore::compensate() then flowWorkspace::estimateLogicle() + flowCore::transform()CATALYST::prepData(..., transform=TRUE, cofactor=5) (arcsinh)Conventional compensation inverts a square spillover matrix (peak-channel subtraction); spectral cytometry solves an OVERDETERMINED least-squares unmix over all detectors, with autofluorescence modeled as an extra "fluorophore." Both correct the population MEAN. Neither removes spreading error - the widening of a negative population in a spillover detector that arises from the Poisson counting statistics of the spilled-in photons (Roederer 2001 Cytometry 45:194; Nguyen 2013 Cytometry A 83:306). Compensation does not INTRODUCE spreading; it makes the pre-existing variance visible by re-centering means. The corollaries are load-bearing: (1) a smeared negative cannot be fixed by tuning the matrix - over-compensating to flatten it is data falsification; (2) spreading is fixed at PANEL DESIGN (the Spillover Spreading Matrix identifies which detector pairs to avoid for co-expressed/dim markers), never downstream; (3) calling spectral unmixing "compensation" is a category error - it is a different, overdetermined model.
| Method | What it does | When to use | Fails when |
|---|---|---|---|
Acquisition-recorded $SPILLOVER | applies the cytometer-computed matrix | trustworthy single-stain setup at acquisition | controls were wrong/missing |
Computed compensation (flowStats::spillover) | estimates spillover from single-stain controls (medians) | conventional flow, controls available | poor/dim/contaminated controls |
| AutoSpill (Roca 2021 Nat Commun 12:2890) | robust-regression matrix + iterative refinement; AF as endogenous dye | high-parameter panels; messy controls | reference implementation/setup unavailable |
| Spectral unmixing (OLS/WLS/Poisson) | least-squares unmix full spectrum vs reference spectra + AF | spectral cytometers (Aurora, ID7000) | wrong/heterogeneous AF; collinear spectra |
| Logicle / biexponential | display + analysis transform, handles negatives | fluorescence flow | wrong w clips the negative population |
| arcsinh | variance-stabilizing transform | CyTOF/mass; computational pipelines | wrong cofactor compresses dim markers |
| log10 | legacy | rarely; strictly positive data | any negative values after compensation |
| Scenario | Recommended | Why |
|---|---|---|
Conventional flow, $SPILLOVER present | apply recorded matrix -> estimateLogicle | trust acquisition controls; logicle handles negatives |
| Conventional flow, no matrix | compute via flowStats::spillover from single-stains (or AutoSpill) | controls drive the matrix; AutoSpill for >12 colors |
| Spectral cytometer | UNMIX (do NOT compensate), then arcsinh at ~150/per-channel (NOT 5) | overdetermined system; spectral data is fluorescence-scale, not ion counts |
| CyTOF / mass | arcsinh cofactor 5; spillover via CATALYST compCytof if needed | metals barely spill (~1-4%), but oxide/impurity is real |
| Dim marker driving a borderline call | test per-channel cofactor (flowVS) | a fixed cofactor can manufacture/erase the population |
Compensation/unmixing is LINEAR and must run on untransformed data; applying it after a nonlinear transform is mathematically invalid. estimateLogicle() must run on ALREADY-COMPENSATED data so the w/a parameters reflect the post-compensation negative spread. Negative values after compensation are expected and meaningful - do NOT clip to zero before transforming (handling negatives is the entire reason logicle/arcsinh exist; log cannot).
Goal: Apply the recorded matrix, or estimate one from single-stain controls.
Approach: compensate() takes a compensation object built from the matrix; flowStats::spillover() estimates from single-stain controls and returns the matrix directly.
library(flowCore)
comp <- compensation(spillover(fcs)[[1]]) # flowCore: flowFrame -> list of keyword matrices
fcs_comp <- compensate(fcs, comp)
library(flowStats)
ctrls <- read.flowSet(list.files('controls', pattern = '\\.fcs$', full.names = TRUE))
comp_matrix <- spillover(ctrls, unstained = 'Unstained.fcs', fsc = 'FSC-A', ssc = 'SSC-A',
patt = '-A$', method = 'median') # flowStats: returns the matrix directly
Goal: Display and analyze compensated fluorescence with negatives handled honestly.
Approach: estimateLogicle() (flowWorkspace) derives w from the data's most-negative events; apply with transform().
library(flowWorkspace)
fluo <- colnames(fcs_comp)[grepl('-A$', colnames(fcs_comp)) & !grepl('FSC|SSC', colnames(fcs_comp))]
lgcl <- estimateLogicle(fcs_comp, channels = fluo) # data-driven w; t=262144, m=4.5, a=0 defaults
fcs_t <- transform(fcs_comp, lgcl)
Goal: Variance-stabilize mass-cytometry counts (or any pipeline feeding clustering).
Approach: asinh(x/cofactor); flowCore's arcsinhTransform is asinh(a + b*x) + c, so set b=1/cofactor. CATALYST prepData defaults cofactor=5.
COFACTOR <- 5 # standard CyTOF cofactor, codified in the CATALYST workflow (Nowicka 2017); ~150 for fluorescence
asinhT <- arcsinhTransform(transformationId = 'asinh', a = 0, b = 1/COFACTOR, c = 0)
fcs_t <- transform(fcs, transformList(marker_channels, asinhT))
# CATALYST path (CyTOF): cofactor=5 default; OVERRIDE for fluorescence/spectral
sce <- CATALYST::prepData(fs, panel, md, transform = TRUE, cofactor = COFACTOR)
Trigger: matrix slope over-estimated from dim controls. Mechanism: subtraction overshoots. Symptom: negative population pulled below zero, "comma" shape. Fix: controls at least as bright as the sample; AutoSpill regression; never hand-tune to flatten spread.
Trigger: fixed w instead of estimateLogicle. Mechanism: linear region too narrow. Symptom: negative population piled on the axis. Fix: estimate w on compensated data.
Trigger: cofactor 5 on fluorescence (or 150 on CyTOF). Mechanism: linear region mismatched to the noise band. Symptom: dim-positive collapses into the negative; clusters don't reproduce. Fix: 5 for CyTOF, ~150 for fluorescence; per-channel via flowVS::estParamFlowVS.
Trigger: treating Aurora data as conventional. Mechanism: subtraction is the wrong model for an overdetermined system. Symptom: residual spread, false positives. Fix: unmix against single-stain reference spectra + unstained AF.
| Threshold | Source | Rationale |
|---|---|---|
| arcsinh cofactor = 5 (mass) | Nowicka 2017 F1000Res 6:748 (CATALYST workflow) | matches CyTOF ion-count near-zero noise band |
| arcsinh cofactor ~150 (fluorescence) | community/CATALYST convention (not a derived optimum) | PMT photon scale is far larger; per-channel flowVS supersedes |
| comp control >= sample brightness | Roederer 2001 Cytometry 45:194 | slope estimated over the widest lever arm; extrapolation amplifies error |
| spreading is intensity-dependent (~sqrt of signal) | Nguyen 2013 Cytometry A 83:306 | SSM is normalized to be gain-independent for panel design |
| Error / symptom | Cause | Solution |
|---|---|---|
compensate() channel mismatch | matrix colnames != FCS channels | align names before compensate |
| all-negative after transform | transform applied before/without compensation | compensate on linear data first |
estimateLogicle not found | called from flowCore | it lives in flowWorkspace |
arcsinhTransform ignores "cofactor" | param is b, not cofactor | set b = 1/cofactor |
まだレビューはありません。使ってみた感想をお寄せください。
概要と使いどころ
生成图文并茂的分析报告/实验报告/质检报告(固定模版)。当用户要求"生成报告/总结报告/实验报告/质检报告/QC报告",或分析流程收尾需要交付报告时使用。报告必须基于真实分析输出,包含固定章节结构、关键指标表格和真实结果图。
日本語の概要は準備中です。原文の説明を表示しています。
Predicts ADMET properties using ADMETlab 3.0 (119 endpoints with uncertainty), ADMET-AI, DeepChem MolNet, and chemprop D-MPNN with explicit handling of OECD QSAR principles, applicability domain assessment, calibration, hERG/CYP/AMES gold-standard endpoints, and PAINS / Lipinski / Ro5 / Veber / BBB druglikeness filters. Use when filtering compounds for drug-likeness, prioritizing leads by predicted safety, or building an in-house ADMET QSAR model.
日本語の概要は準備中です。原文の説明を表示しています。
Trim PCR primers from aligned reads in amplicon-panel BAMs using samtools ampliconclip. Use when processing SARS-CoV-2 ARTIC, hereditary cancer panels, ctDNA hot-spot panels, or any amplicon assay where primer-derived bases would falsely confirm reference at primer footprints.
日本語の概要は準備中です。原文の説明を表示しています。
Filter alignments by flags, mapping quality, and regions using samtools view and pysam. Use when extracting specific reads, removing low-quality alignments, or subsetting to target regions.
日本語の概要は準備中です。原文の説明を表示しています。
Create and use BAI/CSI indices for BAM/CRAM files using samtools and pysam. Use when enabling random access to alignment files or fetching specific genomic regions.
日本語の概要は準備中です。原文の説明を表示しています。
Read, write, and convert multiple sequence alignment files using Biopython Bio.AlignIO. Supports Clustal, PHYLIP, Stockholm, FASTA, Nexus, and other alignment formats for phylogenetics and conservation analysis. Use when reading, writing, or converting alignment file formats.
日本語の概要は準備中です。原文の説明を表示しています。