| QC 1:3s high on one level | Reagent lot, calibrator, or control vial | Repeat QC; inspect open vial dates; check peer QC on same lot |
| QC drift across levels | Calibration curve, lamp/detector, temperature | Recalibrate; maintenance log; vendor service |
| Patient K⁺ 6.5, others normal | Hemolysis | H-index; redraw if clinical discordance |
| Na⁺ 120, normal osmolality | Lipemia VDE or pseudohyponatremia | L-index; direct ISE; confirm serum osmolality indication |
| Glucose ↓30% vs. yesterday | Wrong tube (fluoride/gray), IV fluid contamination, true event | Delta check; specimen type; nursing review |
| PT suddenly ↑ on one patient | Citrate underfill, heparin contamination, factor deficiency | Clot view; mix study; redraw 9:1 fill tube |
| Positive antibody screen, prior negative | Recent transfusion/pregnancy, drug (anti-CD38 → DTT protocol) | History; DTT-treated panel; eluate if needed |
| ABO forward/reverse mismatch | Cold auto, subgroup, leukemia-related weak expression | Repeat on second sample; serologic problem-solving algorithm |
| Blood culture direct AST discordant | Wrong organism ID, polymicrobial, setup >8 h | Purity plate; repeat from isolate per M100 |
| LC-MS ion ratio fail | Ion suppression, column bleed, wrong transition | Re-extract; check SST; investigate matrix lot |
| Autoverification held spike | New middleware rule, AMR boundary, critical delta | Mine held-case log; validate rule per AUTO10 |
| PT/EQA failure | Matrix bias, unit error, transposition | Investigate all failures same analyte; compare to IQC trend |