Guide pump setting adjustments for the MiniMed 780G system based on CareLink data, using Medtronic's 3-step follow-up workflow. Use when a clinician has CareLink data for a 780G patient and wants to know what to change — which setting, by how much, and how to program it. Triggers include: "what do I adjust", "how to fix high TIR", "how to reduce hypos", "adjust ICR/ISF/basal", "SmartGuard settings", "780G pump settings review", or any question about optimising 780G therapy after reviewing a download.

インストール方法を見る

含まれるファイル(1)

  • SKILL.md15.5 KB

SKILL.md(原文)

インストールする前に、エージェントに与えられる指示の中身を確認できます。

MiniMed 780G Pump Adjustment Guide (CareLink-Based)

BEFORE YOU BEGIN

The MiniMed 780G is a hybrid closed-loop system. SmartGuard Auto Mode adjusts basal insulin automatically and delivers Auto Correction boluses, but it cannot compensate for incorrect Bolus Wizard settings. Most therapy problems in Auto Mode stem from meal bolus gaps, not basal errors.

Prerequisites for reliable adjustment decisions:

  • ≥14 days of data (minimum 10 days)
  • SmartGuard use ≥85% (if <85%, the data does not reflect closed-loop performance)
  • Sensor wear ≥85% (gaps = missing data = unreliable patterns)
  • Identify any confounders before adjusting: illness, travel, unusual activity, sensor errors

Golden rule: Change ONE setting at a time. Review in 2 weeks.


THE 3-STEP FOLLOW-UP FRAMEWORK

StepQuestionReport to Use
1. Review therapy goalsAre TIR and Statistics targets met?Assessment & Progress Report
2. Identify the problemWhat pattern is causing out-of-range glucose?Meal Bolus Wizard + Weekly/Daily Review
3. Adjust the settingWhich setting needs changing, and by how much?Device Settings Report

STEP 1 — REVIEW THERAPY GOALS (Assessment & Progress Report)

Time in Range targets (Battelino 2019):

MetricGoalFlag
Time above 250 mg/dL<5%>5% = significant hyperglycaemia
Time above 180 mg/dL<25%>25% = hyperglycaemia problem
TIR 70–180 mg/dL>70%<70% = action needed
Time below 70 mg/dL<4%>4% = hypoglycaemia problem
Time below 54 mg/dL<1%Any value = urgent

Statistics targets:

StatisticGoalIf Not Met
SmartGuard use≥85%Identify and fix SmartGuard Exits
Sensor wear≥85%Counsel on CGM wear
GMI (Glucose Management Indicator)Adults <7.0% / Paeds 7.0–7.5%Surrogate A1C — flag if discordant with lab
CV (Coefficient of Variation)<36%≥36% = unstable glucose, higher hypo risk
Auto Correction % of total bolus<20–25%>35% = meal bolus gap → adjust ICR
Set change interval≤3 days>3.5 days avg = infusion site degradation

Decision after Step 1:

  • All goals met → reinforce current management, review in 3 months
  • Goals not met → proceed to Step 2

STEP 2 — IDENTIFY THE PATTERN (Meal Bolus Wizard + Weekly/Daily Review)

2A. Determine the primary problem

Is it mainly HYPER, mainly HYPO, or both?

Look at the TIR bar on the Assessment & Progress report:

  • Mostly high (orange/yellow) → hyperglycaemia problem → go to Section 3A
  • Mostly low (red) → hypoglycaemia problem → SAFETY FIRST → go to Section 3B
  • Both → address hypoglycaemia first, then hyperglycaemia

2B. Identify WHEN the glucose goes out of range

Use the Percentile Comparison graph (Section 4 of A&P Report):

  • Wide band overnight → overnight basal or SmartGuard target issue
  • Rise after meals → postprandial problem → which meal? Use Meal Bolus Wizard report
  • High fasting in AM → dawn phenomenon or overnight basal

Use the Weekly Review / Daily Review:

  • Post-breakfast rise every day → breakfast ICR too aggressive (ICR g/U too low)
  • Post-lunch rise → lunch ICR
  • Post-dinner rise → dinner ICR; consider high-fat meal and dual wave bolus
  • Overnight highs despite active auto-basal → raise SmartGuard glucose target
  • Auto Correction boluses clustering at same time → systematic meal bolus gap

Use the Meal Bolus Wizard Report:

  • Check Avg SG at Bolus vs SG at 2hr for each meal period
  • Pre-meal glucose rising before bolus → patient not bolusing before meals
  • Post-meal spike then normal at 2h → timing correct, ICR needs adjustment
  • Post-meal spike still high at 2h → ICR definitely too low (not enough insulin)
  • Post-meal lows → ICR too high (too much insulin) or bolus given too early

2C. Check SmartGuard Exits (Section 5 of A&P Report)

Preventable exits (address with patient education):

Exit ReasonCauseAction
BG required for SmartGuardPatient not entering BG when promptedCounsel: enter BG immediately when pump alerts
No SG valuesSensor gap — worn incorrectly or disconnectedCounsel on consistent sensor placement
Sensor ExpiredSensor left in >7 daysCounsel: change sensor on day 6–7
SmartGuard disabled by userPatient manually exiting Auto ModeDiscuss why; address barriers
Prolonged SuspendManual suspend >2hCounsel: resume promptly after suspension

Frequent preventable exits = SmartGuard use <85% = system cannot achieve optimal control. Address exits before adjusting settings.


STEP 3 — ADJUST THE SETTING

PRINCIPLE: ONE CHANGE AT A TIME. Never adjust two settings simultaneously.


3A. HYPERGLYCAEMIA — Setting Adjustments

Problem: High Auto Correction % (>35%) + Postprandial Highs

Diagnosis: Insulin-to-Carb Ratio (ICR) is too low — not enough insulin for meals

Setting to adjust: Carb Ratio (g/U)

  • Lower g/U = more insulin per gram of carb
  • Adjustment: Decrease the carb ratio for the affected meal period by 1–2 g/U
    • Example: Breakfast ICR 13 g/U → try 11 g/U
    • Start with the meal showing the biggest postprandial rise
  • Check: Meal Bolus Wizard report — confirm which meal period is affected (look at which time window shows consistent post-meal rise)

How to program on the pump:

Home screen → Press ⊙ → Settings ⚙ → Delivery Settings → Bolus Wizard Setup → Carb Ratio Select the relevant time period, enter new g/U value, select Done → Save


Problem: Consistent Fasting/Overnight Hyperglycaemia (not postprandial)

Diagnosis: Dawn phenomenon OR overnight SmartGuard target too low for this patient

Option A — Raise SmartGuard glucose target (first-line in Auto Mode)

  • Available targets: typically 100 mg/dL (5.5 mmol/L) or 120 mg/dL (6.7 mmol/L)
  • Adjustment: Change target from 100 to 120 mg/dL if persistent fasting highs despite active auto-basal

How to program on the pump:

Home screen → Press ⊙ → Settings ⚙ → SmartGuard → Target Select new target → Save

Option B — Increase overnight basal rate (if issues persist in Manual Mode or SmartGuard exits frequent overnight)

  • Review the basal pattern in the Device Settings Report
  • Adjustment: Increase the overnight (00:00–06:00) basal rate by 10–20%
    • Example: 0.9 U/hr → try 1.0–1.1 U/hr

How to program on the pump:

Home screen → Press ⊙ → Settings ⚙ → Delivery Settings → Basal Pattern Setup Select active pattern → Options → Edit → Navigate to the 00:00 segment → Adjust U/hr → Done → Save


Problem: Consistent Post-Correction Highs (correction boluses not working well)

Diagnosis: Insulin Sensitivity Factor (ISF) too low — corrections are not strong enough

Setting to adjust: Insulin Sensitivity Factor (ISF / Sensitivity)

  • ISF = how many mg/dL 1 unit of insulin lowers BG
  • Adjustment: If corrections consistently undershoot target, DECREASE ISF by 10–15 mg/dL per U
    • Example: ISF 70 mg/dL/U → try 60 mg/dL/U (more aggressive correction)
  • Warning: Do not reduce ISF if CV ≥36% or frequent lows — this increases hypoglycaemia risk

How to program on the pump:

Home screen → Press ⊙ → Settings ⚙ → Delivery Settings → Bolus Wizard Setup → Sensitivity (ISF) Enter new mmol/L (or mg/dL) per U value → Done → Save


Problem: Glucose Rises 3–4 Days After Set Change

Diagnosis: Infusion site degradation — not an insulin dose problem

Action: Counsel on ≤3-day set rotation. Set a reminder on the pump.

Settings → Reminders → Set Change → Set to 3 days

No insulin dose change needed.


3B. HYPOGLYCAEMIA — Setting Adjustments

⚠️ Address hypoglycaemia before any hyperglycaemia changes. Safety first.

Problem: Post-Meal Lows (<3h after bolus)

Diagnosis: ICR too aggressive (too much insulin for food) OR bolus given too early

Setting to adjust: Carb Ratio (g/U)

  • Adjustment: Increase ICR for the affected meal period by 1–2 g/U
    • Example: Lunch ICR 12 g/U → try 14 g/U (less insulin per gram)
  • Also counsel on bolus timing: bolus at start of meal, not before

How to program: Same path as above → Delivery Settings → Bolus Wizard Setup → Carb Ratio


Problem: Fasting/Between-Meal Lows

Diagnosis: Basal rate too high

Setting to adjust: Basal Pattern

  • Adjustment: Decrease basal rate for the affected time period by 10–20%
    • Example: overnight 0.9 U/hr → try 0.7–0.8 U/hr

How to program: Same path as basal adjustment above → Basal Pattern Setup


Problem: Frequent Low Glucose Alerts + High CV

Diagnosis: ISF too aggressive; or Active Insulin Time too short (insulin stacking)

Option A — Increase ISF (less aggressive correction)

  • Adjustment: Increase ISF by 10–15 mg/dL per U
    • Example: ISF 60 mg/dL/U → try 70–75 mg/dL/U

Option B — Extend Active Insulin Time (AIT)

  • AIT is typically set to 2:00 hours for rapid-acting insulin
  • Adjustment: Extend to 2:30 or 3:00 hours if multiple correction stacking episodes
  • Warning: Extending AIT slows recognition of active insulin → do not exceed 4 hours; confirm with clinical judgment

How to program AIT:

Home screen → Press ⊙ → Settings ⚙ → Delivery Settings → Bolus Wizard Setup → Active Insulin Time Set duration → Save


Problem: Nocturnal Lows (11PM–5AM)

Diagnosis: Overnight basal too high OR evening correction bolus stacking

Adjustment: Reduce overnight basal rate by 10–20%. Also counsel on evening bolus timing — avoid corrections within 2h of bedtime if AIT is 2 hours.


3C. HIGH GLYCAEMIC VARIABILITY (CV ≥36%)

High CV = unstable glucose = increased hypoglycaemia risk regardless of mean glucose.

Common causes:

  • Erratic meal bolus timing (bolusing after eating)
  • Large carb meals without dual/square wave bolus
  • Frequent SmartGuard exits (Manual Mode = no auto corrections)
  • Overcorrecting lows with excessive carbs

Actions:

  1. Reinforce pre-meal bolusing (enter carbs + give bolus BEFORE eating)
  2. Consider dual wave bolus for high-fat/high-protein meals (only in Manual mode)
  3. Reduce SmartGuard exits (see Step 2C)
  4. Counsel on treating lows with precise carb amount (15g rule)
  5. Ensure AIT is correctly set for the insulin used (typically 2 hours for aspart/lispro)

QUICK REFERENCE — FINDING → ADJUSTMENT → PUMP PATH

CareLink FindingProblemSetting to ChangeDirectionPump Path
Auto Correction >35% + post-meal highsICR too lowCarb Ratio ↓ g/UDecrease g/USettings → Delivery → Bolus Wizard Setup → Carb Ratio
Fasting/overnight highsDawn phenomenonSmartGuard Target ↑100 → 120 mg/dLSettings → SmartGuard → Target
Fasting/overnight highs (Manual mode)Overnight basal lowBasal rate ↑+10–20%Settings → Delivery → Basal Pattern Setup
Corrections undershoot targetISF too lowISF ↓ mg/dL per UDecrease mg/dL/USettings → Delivery → Bolus Wizard Setup → Sensitivity
Post-meal lowsICR too highCarb Ratio ↑ g/UIncrease g/USettings → Delivery → Bolus Wizard Setup → Carb Ratio
Between-meal lowsBasal too highBasal rate ↓−10–20%Settings → Delivery → Basal Pattern Setup
Correction bolus stacking / lowsAIT too shortActive Insulin Time ↑+30 minSettings → Delivery → Bolus Wizard Setup → AIT
Site change-related highsInfusion siteSite rotation≤3 daysSettings → Reminders → Set Change
SmartGuard use <85%Preventable exitsEducate patientSee exit type—
CV ≥36%VariabilityBolus timingPre-meal bolusing—

SETTING CHANGE LIMITS AND SAFETY GUARDRAILS

  • Max Bolus: 0–25 U per bolus. Set as indicated by HCP.

    Settings ⚙ → Delivery Settings → Max Basal/Bolus → Max Bolus

  • Max Basal Rate: 0–35 U/hr. Limits Auto Mode as well — if set too low, SmartGuard cannot compensate.

    Settings ⚙ → Delivery Settings → Max Basal/Bolus → Max Basal

  • ICR range: Pump accepts any value; clinical range typically 5–25 g/U

  • ISF range: Pump accepts any value; clinical range typically 1–5 mmol/L per U (18–90 mg/dL per U)

  • BG Target range: Pump-adjustable; SmartGuard target is separate from Bolus Wizard BG target

  • AIT range: 2:00 – 8:00 hours (use 2:00 for most rapid-acting analogues)

Key guardrails:

  • Never adjust more than one setting per visit. Multiple simultaneous changes make attribution impossible.
  • Minimum 3 days of consistent pattern before changing a setting. Do not adjust based on a single outlier day.
  • SmartGuard compensates for small errors — high Auto Correction % often means ICR is the problem, not basal.
  • If SmartGuard use <85%, do not make dose adjustments until exits are addressed and engagement improves.
  • Reinforce that the system is NOT fully automated. Patients must: enter carbs before eating, bolus pre-meal, enter BG when prompted, and follow pump instructions to re-enter SmartGuard after exits.

OUTPUT FORMAT — ADJUSTMENT PLAN

Present adjustments as:

Patient / Date range:

WHAT THE DATA SHOWS:

  • TIR: X% (goal >70%) — [Met / Not Met]
  • Primary problem: [Hyperglycaemia / Hypoglycaemia / Both / Variability]
  • Pattern identified: [e.g. "Consistent post-breakfast rise >250 mg/dL on 8/14 days"]
  • Auto Correction % of total bolus: X% — [Acceptable / High → ICR likely cause]
  • SmartGuard use: X% — [Adequate / Needs attention]

PROPOSED SINGLE ADJUSTMENT:

  • Setting to change: [e.g. Breakfast Carb Ratio]
  • Current value: [e.g. 13 g/U]
  • New value: [e.g. 11 g/U]
  • How to program: [Pump path step by step]
  • Rationale: [One sentence]

PATIENT EDUCATION POINT:

  • [One behavioural focus — e.g. "Pre-meal bolusing at least 5–10 minutes before eating"]

REVIEW IN: [2 weeks — standard after any setting change]


SOURCE

  • Medtronic. MiniMed 780G System with Guardian 4 Sensor: CareLink Reports and How to Follow-Up Patients in 3 Steps (UC202116598a EE ©2021)
  • Medtronic. MiniMed 780G System User Guide (GS3 edition) — Chapter 3: Setting up insulin delivery (pp.67–88); SmartGuard chapter (pp.159–181)
  • Battelino T et al. Clinical Targets for Continuous Glucose Monitoring Data Interpretation. Diabetes Care 2019;42:1593–1603
  • Battelino T et al. Routine use of continuous glucose monitoring in 10,501 people with diabetes mellitus. Diabet Med 2015;2(12):1568–74
  • Danne T et al. International Consensus on Use of Continuous Glucose Monitoring. Diabetes Care 2017;40:1631–1640
  • Bergenstal RM et al. Glucose Management Indicator (GMI). Diabetes Care 2018;41(11):2275–2280

レビュー

まだレビューはありません。使ってみた感想をお寄せください。

同じリポジトリのスキル

概要と使いどころ

Performs corticotropin stimulation test to diagnose adrenal insufficiency when morning cortisol is indeterminate (3-15 µg/dL). Triggers include morning cortisol 3-15 µg/dL requiring further AI testing.

日本語の概要は準備中です。原文の説明を表示しています。

dromlakhani/MD2SKILL132026年10月5日 更新

Administers immediate parenteral injection of 50–100 mg hydrocortisone for suspected adrenal crisis secondary to adrenal insufficiency. Triggers include hypotension, hypoglycemia, or acute illness in a patient with known adrenal insufficiency.

日本語の概要は準備中です。原文の説明を表示しています。

dromlakhani/MD2SKILL132026年10月5日 更新

Clinicians educate adrenal insufficiency patients taking nondexamethasone glucocorticoids who initiate enzyme-inducing antiepileptic drugs about early signs of adrenal insufficiency. Trigger when an AI patient on nondexamethasone GC starts an enzyme-inducing AED such as phenytoin, carbamazepine, or oxcarbazepine.

日本語の概要は準備中です。原文の説明を表示しています。

dromlakhani/MD2SKILL132026年10月5日 更新

In central hypothyroidism patients receiving levothyroxine who initiate enzyme‑inducing antiepileptic drugs, the guideline recommends checking free T4 at least 6 weeks after AED start and increasing the levothyroxine dose if free T4 falls below the target range. Triggers include: central hypothyroidism patient on levothyroxine starting antiepileptic drug such as phenytoin, carbamazepine, or oxcarbazepine.

日本語の概要は準備中です。原文の説明を表示しています。

dromlakhani/MD2SKILL132026年10月5日 更新

Applies serum cortisol cutoffs (<3 μg/dL indicates central adrenal insufficiency; >15 μg/dL likely excludes it) when evaluating morning cortisol for suspected adrenal insufficiency. Triggers include morning cortisol measurement in suspected AI and use when interpreting morning cortisol levels for adrenal insufficiency workup.

日本語の概要は準備中です。原文の説明を表示しています。

dromlakhani/MD2SKILL132026年10月5日 更新

In patients with adrenal insufficiency receiving dexamethasone replacement, the guideline suggests increasing the dexamethasone dose when enzyme-inducing antiepileptic drugs are co-administered. Trigger phrases include "AI patient on dexamethasone starts enzyme-inducing antiepileptic drug" or "AI patient on dexamethasone requiring enzyme-inducing AED".

日本語の概要は準備中です。原文の説明を表示しています。

dromlakhani/MD2SKILL132026年10月5日 更新

dromlakhani のスキルをすべて見る

このスキルの問題を報告する